Français - Nederlands

Vlamfotometer

JenwayBibby Scientific

Industriële Vlamfotometer PFP7 - Jenway

The PFP7 is a low temperature, single channel flame photometer that is designed for the routine determination of sodium, potassium, calcium, barium and lithium concentrations. The flame failure safety system makes these products ideal for use in industrial and educational environments. The use of fine and coarse sensitivity controls allows for accurate measurements each and every time.

Photomètres de flamme_Jenway

Parameter
PFP7
Referentie
500701
Range
0 to 199.9ppm
Limits of detection Na
0.2ppm
K
0.2ppm
Li
0.25ppm
Ca
15ppm
Ba
30ppm
Reproducibility
<1% coefficient of variation for 20 consecutive samples using 10ppm Na set to read 50
Linearity
< 2% error when 3ppm Na/K and 5ppm Li are set to read 100
Stability
< 2% drift over 5min when continuously aspirating 10ppm sample set to 50.0. Specificity Interference from Na/K and Li equal in concentration to test element will be <0.5%
Recorder output
Nominal 1.00V for a reading of 100.0
Electrical supply
90-125V or 190-250V @50/60Hz
Air supply
Moisture and oil free, 6 litre/min @ 14psi
Fuel
Propane, butane, natural gas or LPG
Size (w x d x h)
420 x 360 x 300mm
Weight
8kg
Catalogus Prijs (excl. BTW)
3.994,00 EUR
Levering inclusief : PFP7 industrial flame photometer supplied with Na, K, Ba, Ca and Li filters, connecting hoses and clips, compressor plug and drain trap (230V/50Hz)

 


Klinische Vlamfotometer PFP7/C - Jenway

The PFP7/C is a low temperature, single channel flame photometer that is specifically designed for use in clinical applications for the determination of sodium, potassium and lithium. The in-built lineariser circuitry of the PFP7/C enables readings of both sodium and potassium to be displayed directly in mmol/l. The series 7 diluter is available for use with the PFP7/C to enable serum samples to be diluted efficiently and accurately to the required ratios.

Photomètres de flamme_Jenway

Parameter PFP7/C
Referentie
500801
Range 120 to 160mmol/l Na, 0 to 10.0mmol/l K
Limits of detection Na 0.2ppm
K 0.2ppm
Li 0.25ppm
Reproducibility <1% coefficient of variation for 20 consecutive samples using 10ppm Na set to read 50
Linearity < 2% error when 3ppm Na/K and 5ppm Li are set to read 100
Stability < 2% drift over 5min when continuously aspirating 10ppm sample set to 50.0. Specificity Interference from Na/K and Li equal in concentration to test element will be <0.5%
Recorder output Nominal 1.00V for a reading of 100.0
Electrical supply 90–125V or 190–250V @50/60Hz
Air supply Moisture and oil free, 6 litre/min @ 14psi
Fuel Propane, butane, natural gas or LPG
Size (w x d x h) 420 x 360 x 300mm
Weight 8kg
Catalogus Prijs (excl. BTW)
4.779,00 EUR
Levering inclusief : PFP7/C clinical flame photometer supplied with Na, K and Li filters, connecting hoses and clips, compressor plug and drain trap (230V/50Hz)

 


Vlamfotometer - Sherwood - Niet meer in productie

Photomètre de flamme Modèle 360 - Sherwood_Imlab

The Model 360 Flame Photometer is a LOW COST, single channel, five element, analogue output and digital display Flame Photometer. It has inherited a great deal from our well respected, higher priced models.

The Model 360 Flame Photometer's mixing chamber, burner stem and burner head are taken from our world renowned Model 410 Flame Photometer. Automatic flame optimisation comes from our Dual Channel Flame Photometer Model 420 series of instruments.

Put simply, the Model 360 Flame Photometer is a new product built with tried and tested components where it really matters. It has the best limits of detection available from a low cost instrument and precision comparable to higher cost instruments.

The five element filter stick is easy to remove for cleaning and filter replacement. The mixing chamber, burner head and burner stem come apart for cleaning with no tools required. Electronics and pneumatics come in separate "pods" and both are easily removed from the main instrument chassis; offering a potentially huge saving on shipping costs if a return to base repair or replacement becomes necessary.

The appropriate filter for the element to be analysed is selected using the filter stick and indicated by LEDs on the front panel, which also displays flame status by LED. A display "hold" button allows readings to be captured until recorded - useful if you are unexpectedly called away. For ease of operation, the air regulator is accessible and pressure gauge visible at the front.

A large sample work area is also provided in a spill containment tray which is easy to remove and clean. The instrument is supplied ready for use following installation of the chimney. Connection to the mains power supply is easy; each unit is supplied with a universal switch mode power supply with its own multi-plug adapter kit; no need to worry about voltage selection or making up power cords.